Biomedical Device Publications

Some of the more recent biomedical Case Studies, Solutions Briefs, Technical Articles, and Technical Publications MSI has performed are highlighted below. For older publications please refer to the list at the bottom of this page.

MSI endeavors to make most of our publications available in PDF format for you to download at your convenience. However, due to certain Copyright and distribution restrictions on some publications, we can not provide a PDF file for you to download and therefore request that you fill out the corresponding request form at the bottom of the publication page. We apologize for any inconvenience this may cause.

FEA Model of Spinal Segment Test Fixture

Figure 1. FEA Model of Spinal Segment Test Fixture

Spinal Implant Deployment and In Vivo Performance

A manufacturer of an innovative spinal implant approached MSI for help in assessing the implant’s structural integrity.  Of concern was not only the ability of the implant to withstand the forces during deployment, but also to withstand the in vivo physiologic loading.   The stent-like implant is to function as a mechanical barrier to the escape of nuclear material as would occur during disc herniation.

The spinal disc presents a challenging mechanical environment due to its compliant nature and large deformations.  This intradiscal implant must deform along with the disc as it augments the strength of the annulus.  Another challenge to the strength of the implant is the curved trajectory that the implant must take during insertion to the proper location in the disc.  In order to evaluate the deployment and in vivo stress levels a 3D finite element analysis was performed.  The sliding frictional contact capabilities and the nonlinear material property features of the ANSYS program were employed.

The analysis was performed in several steps.  The disc deformations were first established using an idealized model of the nucleus, annulus and vertebral endplates. Figure 1 simulates the fatigue test fixture comprised of polyurethane components.  Although it differs somewhat from actual spinal anatomy, it does capture the important load conditions for a fatigue assessment.  Sliding contact was modeled between the annulus, nucleus and endplates.  The nucleus was modeled as nearly incompressible.  The spinal segment was then put through a variety of movements including combinations of compression, flexion-extension and lateral bending.  Realistic bulging deformations of the nucleus were ultimately predicted by the model.

The FEA model of the implant, shown in Figure 2, was inserted and then forced to move within the space between the nucleus and annulus as the spinal segment flexed and extended.  Initial models exhibited very high strains primarily due to the insertion.  Subsequent redesigns resulted in acceptable strain levels for both deployment and cyclic in vivo behavior.  Additional analyses were performed with a simulated defect in the annulus, an expected scenario.  Applied pressure would tend to bulge the nucleus out through the annular defect.   The barrier was found to inhibit this bulging.  Development of the implant is ongoing and MSI continues to contribute to the design with these types of simulations.

 Back to Top 

FEA Model and Strain Results for Deployment Figure 2. FEA Model and Strain Results for Deployment

Contact Stress Analysis of a Total Knee Replacement Design

The longevity of an artificial knee replacement depends on many factors. Although great strides have been made in improving the wear performance of ultra-high-molecular-weight polyethylene (UHMWPE), tibial insert wear is still one of the most important failure mechanisms. To improve the wear situation, designs are sought that reduce the contact stress between the femoral and tibial components. Lower contact stress can be achieved by increasing the amount of shape conformance between the femoral and tibial components. Since there is a trade-off between wear performance and kinematic performance it is paramount that the contact stress distribution be well understood for a given design.

Model of knee replacement

MSI has investigated a particular knee replacement design for contact stress. The general surface-to-surface frictional contact algorithms of the ANSYS Multiphysics finite element code were utilized. Higher-order curvilinear elements were employed to model the contacting surfaces. The elastic-plastic material behavior was also incorporated into the UHMWPE tibial insert. The lower surface of the tibial insert was held fixed in space while the femoral component was forced to move according to typical activities including gait cycle motion. Both surface and subsurface stresses in the UHMWPE were calculated and animated throughout the motion. These stress results can also be coupled directly to a wear law so that volumetric wear could be estimated and compared for different designs.

Back to Top

Biomedical Device Publications Order #
Cardiovascular Stent Stress Analysis - A manufacturer of cardiovascular stents developed a new design that required FDA approval before it could be sold in the United States. In order to obtain this approval it had to be shown that the stent could withstand the balloon expansion process without developing any cracks... CS03-0114M
Cement Mantle Stress Analysis for a Total Hip Replacement Design - ... a hip stem manufacturer came to MSI to develop a better understanding of the effect of their hip stem design on the cement mantle stress distribution and to find ways to improve it... CS03-0118M
Contact Stress Analysis of a Total Knee Replacement Design - ...Although great strides have been made in improving the wear performance of ultra-high-molecular-weight polyethylene (UHMWPE), tibial insert wear is still one of the most important failure mechanisms. CS03-0154M
Spinal Implant Deployment and In Vivo Performance - A manufacturer of an innovative spinal implant approached MSI for help in assessing the implant’s structural integrity... CS03-0153M
Spinal-Implant Fatigue Stress Analysis - A manufacturer of spinal implant systems had developed a new external-fixation clamping device to the stage that it was ready to undergo required compression-fatigue testing... CS04-0155M
 